Description
Thermocouple data logger tracks temperature input from types J, K, T, E, R, S, B or N thermocouples
Micro T/C Temperature Logger provides a versatile option for a broad range of temperature recording applications from cryogenics to glass with the capability to store 16,383 temperature readings per channel using a wide variety of thermocouples. The Logger's compact size and large storage capacity make it ideal to record temperature data during processing or in locations with difficult access.
The Micro T/C accepts thermocouple probe types J, K, T, E, R, S, B or N to record time and date stamped temperature readings from -454 °F to 2372 °F (-270 °C to 1300 °C) and readings in the range of -40 °F to 176 °F (-40 °C to 80 °C) with its internal sensor.
The Micro T/C automatically performs cold-junction compensation and linearization and provides the logged temperature data from its internal sensor as a second temperature channel. The lithium battery-powered temperature logger records the time and date stamped temperature readings in °C, °F, °K or °R, at user-definable recording intervals from two seconds to 12 hours.
Before recording, the software included in the low-cost MLSP Support Package (see details below) is used to identify the thermocouple type, select a reading rate and assign a User ID to each recording session.
The 8540-00 is a three foot long PTFE/Plastisol insulated Type K thermocouple with mini-connector. It is suitable for benign environments to measure temperatures as cold as dry ice up to 180 °F. Other thermocouples with mini-connectors are available.
Specs
Micro T/C Specifications | |
---|---|
Temperature Range | -40 °C to 80 °C (-40 °F to 176 °F) |
Temperature Accuracy | ± 0.5 °C (0 °C to 50 °C) ±1.0 °C (<0 °C and>50 °C) |
Response Time | 500 milliseconds |
Resolution | 0.1 °C |
Time Accuracy | ±100ppm at 68 °F (20 °C) ± 1 min/month |
Storage Capacity | 16,383 readings maximum per channel |
Power Source | Replaceable 3.6 V lithium battery (included) |
Recording Rates | 2, 5, 10, 15 and 30 seconds 1, 5, 10, 15 and 30 minutes 1, 2, 4, 8, and 12 hours |
Data Format | Date and time stamped in °C, °F, °K or °R, mV |
Dimensions | 0.6" x 1.4" x 2.15" (15 mm x 35 mm x 54 mm) |
Weight | 1.0 oz (25.5 grams) |
Visual Indicator | LED flashes at customer selected reading rate |
Software | Windows® based software for complete control |
Micro T/C Ranges and Accuracy* | ||
---|---|---|
Type | Ranges | Accuracy |
J | -210 °C to 760 °C (-346 °F to 1,400 °F) | 1.8 °C (1.0 °F) |
T | -200 °C to 350 °C (-328 °F to 662 °F) | 1.5 °C (2.7 °F) |
K | -200 °C to 1,250 °C (-328 °F to 2,282 °F) | 1.2 °C (2.0 °F) |
E | -200 °C to 1,000 °C (-328 °F to 1,832 °F) | 1.3 °C (2.1 °F) |
R | 0 °C to 1,450 °C (32 °F to 2,642 °F) | 2.4 °C (4.3 °F) |
S | 0 °C to 1,450 °C (32 °F to 2,642 °F) | 2.4 °C (4.3 °F) |
B | 0 °C to 1,700 °C (32 °F to 3,092 °F) | 2.7 °C (4.9 °F) |
N | -270 °C to 1,300 °C (-454 °F to 2,372 °F) | 1.0 °C (1.8 °F) |
*Accuracy excluding thermocouple error |
Data
Data Analysis Software For Micro Series Loggers
The Telatemp MLSP Support Package includes everything you need to set up, start and download data from any number of data loggers in the Micro Series using a PC. To get started, the user connects the Interface Cable to the logger and to the PC's serial port, inputs the device ID description and starts the data collection through the easy-to-use software. The software is compatible with Windows 95 or newer and configures itself appropriately.
After data collection, the logger connects via the Interface Cable to the PC to list and/or graph the data or, export the data as a text file or as a .CSV file for import into Microsoft Excel and similar programs. Special features include "real time display" which allows the user's computer to emulate a strip chart recorder by connecting the logger to the PC and monitoring and recording simultaneously. All data can be annotated.
Also provided is statistical analysis for the whole data set or any "zoomed" subset of the data. Some of the statistics provided are minimum, maximum, arithmetic mean, standard deviation and mean kinetic temperature.
Part No. | Description | Quantity In Stock |
Price EA |
Order |
---|---|---|---|---|
MLTC | MICRO T/C TEMPERATURE LOGGER | DISCONTINUED | --- | --- |
MLSP | MLSP SOFTWARE SUPPORT PACKAGE INLCUDES INTERFACE CABLE, SOFTWARE AND OPERATION MANUAL | DISCONTINUED | --- | --- |
854000 | TEFLON/PLASTICOL 3' TYPE K THERMOCOUPLE | DISCONTINUED | --- | --- |
854001 | TEFLON/PLASTICOL 6' TYPE K THERMOCOUPLE | DISCONTINUED | --- | --- |
854002 | TEFLON/PLASTICOL 9' TYPE K THERMOCOUPLE | DISCONTINUED | --- | --- |
MLSPUSB | MLSP SOFTWARE SUPPORT PACKAGE INCLUDES USB INTERFACE CABLE, SOFTWARE AND OPERATION; MANUAL | 0†Standard lead-time for non-stock items is approximately two weeks depending on the item. | $155.00 | |
* Call for availability. † Standard lead-time for non-stock items is approximately two weeks depending on the item. | ||||
Ordering information. International (non-US) prices available through your local dealer. |