Description
Compact .5 °C accuracy 16,000 point datalogger records temperature vs. time of temperature sensitive products in transit - recovers data on a PC
The Telatemp PCT-2000 Transit Thermometer is a high performance, tamperproof .5 °C accuracy real-time vs. temperature recording system designed to cost effectively protect the quality and safety of critical temperature sensitive goods in transit.
The Telatemp PC Transit Thermometer system consists of: A compact, battery operated, self contained data logger module that travels with a shipment and the support package: a data recovery interface with 9-pin serial connector that cradles the data logger to transfer data to an IBM PC or compatible, Telatemp Transit Software for data logger set up, data capture, display, analysis, graphing and printing and the key ring magnetic starter.
Data Logger Module
At the heart of the PC Transit Thermometer system is the .5 °C accuracy solid state PCT-2000-DLM Data Logger Module which senses and records up to 16,000 temp vs. time points from -40 °C to 70 °C. Timing accuracy is 40 sec./month. The rate of data sampled can be programmed for intervals from one to 255 minutes. Recordings are indexed to "real time", and can be read in local time, sender's time, or elapsed time.
Rugged, Water-resistant
Housed in a rugged polycarbonate water-resistant case, the compact 2.5 oz, 3" dia. x .75" data logger module can operate under adverse environmental transit conditions with 100% reliability - trip after trip.
Permanent Memory
The non-volatile "permanent" memory assures that all recordings will be retained in memory, even if the 3.6V lithium battery becomes depleted or removed.
Start Recording
To start recording a trip with the data logger is simple. It can be pre-programmed to start at a particular time. Or it can be started by the shipper with a convenient Key Ring Magnetic Starter that is swiped across the logger's back side. Beeping tones tell when the logger is operating. The logger then travels with the load to record temp/time trip activity.

Data Recovery Interface
At the destination, the trip data can be quickly retrieved by placing the logger into its Data Recovery Interface PC-2000-DRI and loading the stored information into an IBM PC or compatible using the Telatemp Transit Software. Recorded data is totally protected against unauthorized use, tampering or duplication with a secure password programmed in software.
Versatile Transit Software
The menu driven Telatemp Transit Software is easy to use. It sets up the parameters for temp/time recordings in the Data Logger Module, including the interval between recordings and the cargo's minimum and maximum safe temperature ranges. The trip number is automatically assigned for each use.
Reporting Options
Recorded data can be displayed and printed in List and Graph modes in °C or °F. A variety of options are available for formatting, printing, graphing and saving the data. Reports include a serial number, the trip number, sampling frequency, a summary of all measurements outside of the cargo's safe minimum and maximum temperatures. Also, time-stamped temperatures for an entire trip's recordings with out of spec temperatures noted with a * are listed.
The average temperature of the trip or any portion of the trip can be shown, along with the total number of logs and the number of readings outside of the safe temperature band.
Trip data can be saved to disk in the Telatemp compact format to be queried at any time by the Transit Software, or as an ASCII file or .SLK file for export to popular spreadsheets, databases and word processing programs. Data can be output on popular printers.
Specs
PC Transit Specifications | |
---|---|
Temperature Accuracy | ± 0.5 °C (1.0 °F) |
Temperature Resolution | <1.1 °C from -40 °C to -20 °C <0.5 °C from -20 °C to 40 °C <1.1 °C from 40 °C to 70 °C |
Sample Rate | Programmable - intervals of one minute to 255 minutes |
Data Capacity | 16,000 data points nominal |
Measuring Range | -40 °C to 70 °C (-40 °F to 158 °F) |
Time Accuracy | 40 seconds per month. Logs in real time |
Storage Temperature | Recommended less than 27 °C (80 °F) |
Power Source | Replaceable 3.6 V lithium battery (included) 2 year life |
Starting Device | On site activation by magnet contact or user programamable for activation at specific date and time |
Function Check | Beeps on start up to confirm activation and beeps while logging |
Software | Menu driven format using Windows |
Data Recovery | Logger mates with Data Recovery Interface for connection to IBM compatible PC |
Case Material | Logger and Interface are made of polycarbonate with plated metal contacts |
Dimensions | Diameter: 3.00" (7.62 cm) excluding mounting lug Depth: 0.725" (1.84 cm) Mounting Hole Diameter: 0.32" (0.80 cm) |
Unit Weight | 2.5 oz (70 grams) |
Part No. | Description | Quantity In Stock |
Price EA |
Order |
---|---|---|---|---|
PCT2000DLM | DATA LOGGER MODULE ONE INTERNAL TEMP SENSOR, 16000 POINTS | DISCONTINUED | --- | --- |
PCTIHDLM | DATA LOGGER MODULE ONE INTERNAL TEMP SENSOR AND ONE INTERNAL HUMIDITY SENSOR | DISCONTINUED | --- | --- |
PC2000LSPW | LOGGER SUPPORT PACKAGE INCLUDES DATA RECORVERY INTERFACE, WINDOWS TRANSIT SOFTWARE, KEY; RING MAGNETIC STARTER | DISCONTINUED | --- | --- |
PC2000DRI | DATA RECORVERY INTERFACE INLCUDES 9-PIN SERIAL PORT CONNECTOR FOR IBM AND COMPATIBLES | DISCONTINUED | --- | --- |
PC2000TSWW | TRANSIT SOFTWARE-WINDOWS | DISCONTINUED | --- | --- |
PC2000KRS | KEY RING MAGNETIC STARTER | DISCONTINUED | --- | --- |
* Call for availability. | ||||
![]() |