![]() |
89 | Last Update: 09/12/00 |
Article by: John J. Manura and David J. Manura
Scientific Instrument Services, Inc.,
1027 Old York Road, Ringoes, NJ 08551
Introduction:
Figure 1 – AutoProbeÔ attached to the Finnigan TraceÔ MS from ThermoQuest
This study describes the application of a new automated direct exposure MS probe (AutoProbeÔ), which overcomes the limitations of manual probe techniques. The AutoProbe can perform an entire MS sample analysis reproducibly, quantitatively, and unattended in under 3.0 minutes. In a typical sample analysis, the AutoProbe precisely places the sample on the direct expose probe (DEP) wire, reproducibly introduces the sample into the mass spectrometer, and reproducibly heats the wire according to a current program so that the sample may be analyzed by the MS. Variability in sample handling and analysis are greatly minimized, resulting in reproducible and quantitative analysis of samples with both EI scanning and SIM mass spectrometer techniques.
The
AutoProbeÔ System from Scientific Instrument Services
(SIS), Inc. is attached to the Finnigan TraceÔ MS system from
ThermoQuest. The AutoProbe System is designed to inject samples directly into the MS
source. The AutoProbeÔ system consists of a CTC Autosampler attached to
an auto injecting direct exposure probe (DEP) as shown below (Figure 2). (The
GC in Figure # 1 is not used in this configuration.) The
Trace MS is controlled and operated by the ThermoQuest XcaliburÔ software. The
TraceÔ MS system is a quadrupole MS system fitted with a dual
EI/CI source. The MS has a mass
range of 1000 daltons. For these
studies the MS was operated in the EI state under both the mass scan and the SIM
(single ion monitoring) modes.
Figure
2 – Components of the AutoProbe System
AutoProbe System Components
The Direct Exposure Probe (DEP) tip uses a platinum wire coil for the analysis of samples. Samples for analysis are dissolved in a suitable solvent and placed into the Autosampler tray by the user. The Autosampler then injects 0.125 to 1.0 ul of the prepared sample solution onto a small platinum wire loop on the tip of the probe.
Figure 3 – Loading sample onto the DEP Filament Wire
The
AutoProbe is designed to interoperate with the ThermoQuest XcaliburÔ MS Software and the Finnigan TraceÔ MS. The
AutoProbe is fully integrated with the Xcalibur software operation, including
system configuration, methods setup, sequences, sample runs, and data storage,
resulting in a seamless probe-mass spectrometer operation package.
In addition the technique has proven to be easily used in an Open Access
environment.
The
standard sequence of operation of the AutoProbe System is as follows.
A series of organic compounds with molecular weights ranging from 200 to 400 amu were each dissolved in a solvent (either methanol or chloroform), and successive dilutions were made from 500 ng/ul down to less than 1.0 ng/ul . About 25 different concentrations of each of the compounds were prepared. The following is a typical set of standards prepared: 500, 450, 400, 350, 300, 250, 200, 180, 160, 140, 120, 100, 80, 60, 40, 20, 16, 12, 8.0, 4.0, 2.0, 1.0, 0.5, 0.2, 0.1 (nanograms per microliter). About 500 ul aliquots of each of the dilution were placed in individual 1.8 ml sample vials, which were then placed into the sampling tray on the AutoProbe's CTC Autosampler. Each sample was analyzed in triplicate for each method, and in a few cases a sample was analyzed five sequential times to determine the reproducibility of the analysis. In addition a blank methanol sample was analyzed between each different concentration.
Samples
of cholestane (MW = 372 ) were
prepared in chloroform to conduct a number of studies on the automated DEP
technique. Several parameters such
as sensitivity, reproducibility, cross contamination, memory effects, and
quantitation were studied with the cholestane standards.
Reproducibility - To determine reproducibility of the AutoProbe technique, a sample of cholestane at a concentration of 125 ng/ul in chloroform was prepared. Then 0.5 ul of this sample (62.5 ng) was repeatedly injected and analyzed via the AutoProbe on the MS. The DEP filament was heated with 900 mA of current for 30 seconds during analysis and subsequently baked out at 1000 mA for 15 seconds in the bakeout cleaning step. This sample analysis was repeated 158 times. The mass spec was scanned over the mass range of 100 to 400 daltons and the total ion signal was collected. The sample data was analyzed by extracting the 217 ion signal of the cholestane and integrating the area of this ion. The peak area at mass 217 for each sample of the 158 samples was plotted and is shown in the chart below. The reproducibility of peak areas for all the samples was within 10% of each other. This study demonstrated that the AutoProbe DEP technique is very reproducible. This data also suggested that the automated DEP technique might be quantitative, which will be demonstrated in the subsequent studies.
Figure
4 – Reproducibility study of Cholestane samples
Historically
the direct probe techniques have not been known to be reproducible.
It is quite difficult to accurately and reproducibility apply a sample to
a DEP probe tip. In addition the DEP probe tip wire is often touched, the
coils distorted or the coils shorted out when samples are injected onto the DEP
filament wire. Changing the geometry of the DEP filament coil changes its
heating patterns and therefore minimizes the chance for reproducibility.
The AutoProbe eliminates these problems.
"Memory Effects" - To test "memory effects" with the DEP technique, cholestane samples were prepared in chloroform at a concentration of 500 ng/ul. An injection of 0.5 ul of this sample (total 250 ng of cholestane) was injected onto the DEP probe tip and analyzed as described previously. This analysis was completed five consecutive times within a 15 minute time span (3.0 minutes per sample). After each sample run, the DEP wire was baked out at 1000 mA for 15 seconds. After the fifth cholestane sample was analyzed, 0.5 ul of methanol was injected onto the DEP probe tip and analyzed. Both plots shown below are at the same intensity scale (Figure 5). The cholestane sample produced a strong total ion chromatogram. The blank sample produced a flat baseline that was only observed at a scale expansion of 1000 times (three decades of expansion) the cholestane sample.
Figure
5 - Background check after running 5 consecutive strong samples.
The
second chart below (Figure 6) shows the mass spec data at 0.18 minutes for the
250 ng of cholestane and the blank methanol sample from the data in the total
ion chromatograms above. The scale
on the blank sample is at 1000 times the scale expansion of the 250 ng of
cholestane sample mass spectrum. No
peaks were detected in the blank sample that could be attributed to the
cholestane.
Figure
6 - Background mass spec after running 5 consecutive strong samples
The
results above demonstrate that the DEP technique does not suffer from
"memory effect" or "cross contamination" problems.
The high temperature cleaning step removes all traces of any of the
previous samples. It is recommended
that the high temperature cleaning step be used in all analysis.
Running a blank sample between samples is optional.
Quantitation
- Cholestane samples were prepared in chloroform in concentrations from 500 ng/ul
down to 1.0 ng/ul as described above. The samples were analyzed with the
AutoProbe as previously described. The
mass spectrometer was operated in the EI mode (70 eV), operated in the total ion
scan mode and scanned from mass 100 to 400 daltons (2.9 scans/second).
Each of the prepared sample dilutions was analyzed five times to
demonstrate that the automated DEP technique was quantitative and to show its
reproducibility. After each set of
five samples a blank methanol sample was run.
Figure 7 - Calibration Curve for the Quantitation of Cholestane
The
results verified that the automated DEP technique is not only reproducible, but
it is quantitative. Quantitation
with a direct probe to our knowledge has not been achieved in the past.
The reproducibility and quantitation demonstrated by the AutoProbe
technique are due to the automation of the AutoProbe system with the exact same
conditions being applied to each sample and to the unchanging geometry of the
DEP filament. Since the filament coil is never touched or moved, the sample
is injected onto the DEP filament coil in exactly the same position of the
DEP coil for every sample. This
provides for many new applications of the DEP technique to enable the fast
quantitation of analytes.
Slow
Ramp Study
– A series of additional studies were conducted by heating the DEP filament
with a programmed ramp from 0 to 900 mA at various ramp rates of 50, 100, and
250 mA per second to study the effect of a slower temperature ramps on the total
ion chromatogram peak shape and sensitivity.
The resulting data was still linear, but the slower ramp rates produced
lower total ion peak heights and therefore reduced the sensitivity of the
technique. Also due to the lower
peaks heights, increased peak splitting and increased tailing, the
reproducibility of the integration of the peak data deteriorated.
Another steroid compound, prednisone (MW = 358), was prepared as described above at concentrations from 500 to 1.0 ng/ul in methanol and analyzed by ballistically heating the AutoProbe tip with 900 mA of current. The major EI ions for prednisone are 299 and 358. Two different MS methods were used for the analysis. In the first MS method, the samples were analyzed in the total ion scan mode over the mass range from 155 to 362 daltons (4.5 scans/second). In the second MS method the MS was operated in the SIM (single ion monitoring) mode and data was collected at 299 and 358 daltons (15 points/second) with a mass variation of +/- 0.5 daltons. Data for the 299 and 358 ions in the mass spectra of prednisone were integrated from each of the MS methods and the data is shown below (Figure 8)

Figure
8 – Quantitation of Prednisone – Total Scan (left) and SIM mode (right)
The plots for both the total scan and SIM methods and for both the 299 and 358 ions are linear. The SIM does have the advantage of increased sensitivity as expected. The limit of sensitivity for the total scan analysis of prednisone is about 2 nanograms, whereas the sensitivity for the SIM technique is less than 0.5 nanogram.
Next
a series of alkaloids were studied to further verify the quantitative
capabilities of the AutoProbe technique. The
first of these was ibuprofen (MW = 358).
Ibuprofen dilutions from 500 to 1 ng were prepared in methanol as
described previously. The AutoProbe
was heated ballistically with 900 mA of current.
The major EI ions for ibuprofen are 206 and 358.
The MS was operated in the EI total scan mode (202 to 360 daltons – 5.5
scans/second) and the SIM mode (ion 206 and 358 – 15 points/second).
The area of the ibuprofen major 206 ion was integrated and plotted as
shown below (Figure 9).
Figure
9 – Quantitation of Ibuprofen – Total Ion Scan (lower plot) and SIM mode
(upper plot)
The data was linear over the concentration range of 200 to 0.5 nanograms. The sensitivity of the SIM technique greater than the total ion scan method. The SIM technique would have been detectable down to at least 0.1 nanograms had these lower concentrations of samples been prepared.
The
second alkaloid studied was diazepam. Diazepam
dilutions from 500 to 0.4 nanograms were prepared in methanol.
Again, two MS methods were used for the analysis.
In the total MS scan, the MS was scanned over the MS range of 171 to 311
daltons (6 scans/second), and in the SIM mode data was collected at masses 256
and 283 (15 points/second). The
areas of the 256 and 283 ions were integrated and plotted as shown below (Figure
10).

Figure
10 – Quantitation of Diazepam – Total Ion Scan (lower plots) and SIM mode
(upper plots)
A
number of samples have been studied via the AutoProbe DEP filament technique to
determine the reproducibility of the AutoProbe system and to determine the
accuracy of quantitative methods utilizing AutoProbe. Samples were analyzed from
250 nanograms down to 0.1 nanograms in both the MS total ion scan mode, as well
as the MS SIM mode. The resulting
plots of sample size versus total areas of an abundant ion in the MS
chromatogram were linear over more than two decades of sensitivity.
The results proved that the automated DEP technique is not only
reproducible but also quantitative. In
addition "memory effect" problems do not occur with this technique when the
high temperature bakeout is used to clean the DEP filament wire after a sample
analysis.
Quantitation
with a direct probe has not been reported in the past.
This reproducibility and quantitation capability of the AutoProbe
technique is due to the automation of the system with the exact same conditions
being applied to each sample and to the unchanging geometry of the DEP filament.
Since the filament coil is never touched or moved, the sample is injected
onto the DEP filament coil in exactly the same point for every sample.
This opens up many new applications of the DEP technique for the fast
quantitation of analytes.
The
AutoProbe technique has been proven to be a quantitative MS injection technique
that can reproducibly and quantitatively analyze higher molecular weight samples
in about three minutes per sample. The
technique will be invaluable in the analysis of samples not only in a quality
control laboratory but also in a laboratory where high volumes of samples need
to be analyzed and quantitated in a short period of time.
![]() Yttria coated filament at start |
![]() Yttria coated filament after 16,000 cycles |
New Pumps Added to the List of LOW COST S.I.S. Rotary Vane
Vacuum Pumps - Save 35 - 50% off the price of comparable vacuum pumps.
Adixen (Alcatel) ACP Series Compact Dry Roughing Pumps - a frictionless pumping module, optimized for operation without internal lubricant.
- Direct Replacement for the Shimadzu QP5000 Filament.
PCI Analytics Products - product solutions for all Analytical Instrumentation
SilTite FingerTite, the next generation GC capillary ferrule system without the use of any tools.
(25% discount until Oct 31)
Direct Replacement for the Heated Capillary in the Thermo-Fisher Scientific TSQ & LCQ Mass Spectrometers now available
- Ideal for use in analytical instrumentation
- Will improve your laboratory workflow and reporting confidence in many ways
DigiVac Bench Top Vacuum Gauges
- Vacuum instrumentation with everything you need to go to work
Imtakt HPLC Columns
-offer a wide variety of unique stationary phases in their HPLC
columns
Programmable syringe pumps for controlled flow. Model NE-300 priced at only $275.00
NEW: Wiley 9th Edition - 660,000-850,000
spectra (a 65% increase).
NEW: Agilent 5973/75 Replacement Source
Heater and Temperature Sensor
NEW: SIS Direct Replacement
Filament for the Amatek/Dycor RGA Leak Detectors and Quadrupoles
(p/n 95381VE), which tests have shown increase service life.
NEW: SIS
Replacement Filter Elements For Edwards Oil Mist Eliminators -
Save up to 25% off the price of comparable Edwards replacement
elements.
Identify mass spectra via library
search with the fully-evaluated, most widely used mass spectral
reference library, compiled by the National Institute of Standards
and Technology (NIST).
View and capture images or video at
20x/400x power on a PC computer via attached USB cable.