Technical Data
Number of Stages
2
Displacement (swept volume) 60 Hz
1.3 ft
3
min
-1
(2.2 m
3
h
-1
)
Speed (Pneurop 6602) 60 Hz
1.2 ft
3
min
-1
(2.0 m
3
h
-1
)
Ultimate vacuum (total pressure)
gas ballast closed
1.1x 10
-3
torr(2.5 x 10
-3
mbar)
with gas ballast
1.9 x 10
-2
torr(2.5 x 10
-2
mbar)
Inlet connection
NW10 flange
Outlet connection
Nozzle 11 mm external
diameter removable from
hole tapped 3/8" BSP
Maximum allowed pressure at outlet
7 psig (0.5 bar gauge)
Maximum inlet pressure for water vapor
11 torr (15 mbar)
Maximum water vapor pumping rate
0.16 kgh
-1
Motor power 60 Hz
1/4 hp (0.16 kW)
Noise
54dB(A)
Oil Capacity maximum
0.28 liter
minimum
0.20 liter
Weight
22 lbs (10 kg)
Technical Data
Displacement (swept volume) 60 Hz
2.6 ft
3
min
-1
(4.5 m
3
h
-1
)
Speed (Pneurop 6602) 60 Hz
2.3 ft
3
min
-1
(3.9 m
3
h
-1
)
Ultimate vacuum (total pressure)
7.6x 10
-5
torr(1 x 10
-1
mbar)
Inlet connection
1/2" id hose adapter
Outlet connection
1/2" id hose adapter
Maximum allowed pressure at outlet
14 psig (1 bar gauge)
Maximum allowed inlet and gas ballast pressure
7 psig (0.5 bar gauge)
Maximum inlet pressure for water vapor
30 torr (40 mbar)
Maximum water vapor pumping rate
0..95 kgh
-1
Mist Filter oil trap efficiency
99.999%
Motor power 60 Hz
0.30 kW
Operating temperature range
12-40°C
Noise
50dB(A)
Oil Capacity maximum
0.7 liter
minimum
0.42 liter
Weight
47.3 lbs (21.6 kg)
All dimensions in millimeters
All dimensions in millimeters
42.5 42.5
142
110
320
140
65
124
154
A
B
171
226
140.5
11
69
Ø12.7
127
120
230
156
430
158
450
4 x
9
This miniature two stage pump features an alternative inlet con-
nection position at the side of the pump so that overall installation
height can be reduced to a minimum when required. The interstage
version features an additional pumping port to provide a rough vacu-
um for auxiliary duties. This pump is widely used on mass spectrom-
eters including the Agilent 5971, 5972, 5973, and 5975
The E-LAB 2 pump includes all the components needed for labo-
ratory applications such as wet chemistry and sample preparation.
Based on BOC Edwards proven past experience with vacuum pumps
and developments in accessories, the E-LAB 2 pump allows continu-
ous running, even in the presence of water and solvents. The pump is
supplied with: outlet mist filter, oil return kit, oil change tag, nozzles,
elbow, all fittings, 1 liter of oil, instruction manual and and electrical
supply cable. Assembly is necessary.
Part No.
Description
BOC Edwards E2M1.5 Pump
110-120V, 60Hz
BOC Edwards E2M1.5 Pump
220-240V, 60Hz
BOC Edwards E2M1.5 Pump 110V
with cord for Agilent MSD
EMF3 Oil Mist Filter
Clean and Overhaul Kit
Blade Kit
Part No.
Description
E-LAB2 pump, 110-120V
E-LAB2 pump, 220-240V
Clean and overhaul kit
Blade kit
tip...