Thermocouple Vacuum Gauge
F28
Phone: 908-788-5550
www.sisweb.com
The Varian Thermocouple Vacuum Gauge Tube Model 531
Filament Current
165 mA (gives 14 mV across the thermocouple at pressures below 10 -3 Torr)
Response Time
3 sec.
(10 -3
to 1.0 Torr)
Range
1 x 10 -3 to 2 Torr
Output
0 to14mV
Heater Resistance
1.4 ohms (approx.)
Dimensions
1.21" O.D. x 3" long
Connections
May be soldered into a system, used with1/2" '0' ring compression seal or
threaded on 1/8" IPS female thread.
Cable Connection
Octal Base (std.)
Casing
Nickel-plated mild steel
Pins
Nickel-iron alloy (Ni 48%, Fe 52%)
Volume / Weight
12 cc / 2 oz.
General Description
The Model 531 Thermocouple Vacuum
Gauge provides continuous and accurate
indication of total pressure from 1 x 10
-3
to 2
Torr absolute. Developed to combine
rugged dependability and outstanding per-
formance, the 531 offers fast response and
low power drain.
The 300 Series Digital Thermocouple Vacuum Gauge Controls are used for repeatable,
stable pressure measurement in the 10
-3
Torr to 2 Torr range. These controls are fully
temperature compensated by use of a thermistor placed in the gauge cable socket. The
thermal response of the thermistor is equal to, but in the opposite direction from the
thermocouple characteristics, thereby compensating for ambient temperature changes.
The 302
Part #302—Click for details
DISCONTINUED
Unit of Measure:
EA
Est. Ship Weight=4 lbs
Control is compact and lightweight, with three mounting studs to facilitate
panel mounting in a minimum of space. The 302A
Part #302A—Click for details
DISCONTINUED
Unit of Measure:
EA
Est. Ship Weight=4 lbs
is mounted in a portable enclosure.
The 302
Part #302—Click for details
DISCONTINUED
Unit of Measure:
EA
Est. Ship Weight=4 lbs
and 302A
Part #302A—Click for details
DISCONTINUED
Unit of Measure:
EA
Est. Ship Weight=4 lbs
have easy-to-read meter scales, recorder output, and include a 10
ft. gauge cable and one 531 thermocouple gauge tube. It also includes a 6-ft. power
cord. All necessary cables are included. Also available in 220V.
The Varian 300 Series Digital Thermocouple
Vacuum Gauge Controls
Specifications
Principles of Operation
T
hermocouple gauges measure pressure in terms of the amount of
heat that gas will conduct away from a hot wire. The sensing
element is a combination heater and thermocouple. Cross-shaped, it
is formed by joining two short resistance wires of different compositions, a
method of construction which gives excellent shock resistance and yet
provides the minimum mass required for
fast response. With a fixed heater
power, the temperature of the junction, which depends on the conductivity of the
gas, is indicated by the voltage output of the thermocouple. As the
pressure drops, the thermocouple gets hotter and its output increases. Readout
is accomplished on a millivolt meter calibrated to indicate pressure.
Part No.
Description
F0472301
Part #F0472301—Click for details
Current Price: $160.00
Unit of Measure:
EA
Est. Ship Weight=0.25 lbs
Quantity in Stock=2
Varian 531 Vacuum Gauge Tube
Part No. *
Description
302
Part #302—Click for details
DISCONTINUED
Unit of Measure:
EA
Est. Ship Weight=4 lbs
Panel Mount Vacuum; Gauge Control, DIGITAL
302A
Part #302A—Click for details
DISCONTINUED
Unit of Measure:
EA
Est. Ship Weight=4 lbs
Vacuum Gauge Controller, Mounted, DIGITAL
* For 220V Controls add “220” to each part number (i.e. - xxx-220)
302A
Part #302A—Click for details
DISCONTINUED
Unit of Measure:
EA
Est. Ship Weight=4 lbs
3.03
(77)
0.5
(12.7)
1.21
(30.7)
Dia
Hex.
.56
(14.2)
A.P.
Dimensions
Inches
(Millimeters)
0.5
(12.7)
0.4
(10.2)
.50
(12.7)
Dia
.41
(10.4)
.12
(3.05)
- Table of Contents
(Print Catalog)Table of Contents of the Scientific Instrument Services general catalog volume 36. - Print Catalog F. Vacuum System Supplies SIS Vol. 36 catalog Section F - Vacuum System Supplies.
- Ionization Gauge Tubes
(Catalog F27) - Thermocouple Vacuum Gauge
(Catalog F28) - Ionization Gauge Tubes
(Catalog F29)