Probe Application Note No. 2
Thermal Desorption System Problem Solving - March 1994
by Christopher Baker and John J. Manura,
Scientific Instrument Services, Inc.,
Presented at PITTCON Meeting, Chicago, IL. MARCH 1994
Introduction
The use of a direct insertion probe is a common method for the introduction of samples into the mass spectrometer. The technique is preferred due to the speed of analysis of solid matrix samples. Direct probe samples can be analyzed in as little as 2 minutes as compared to GC samples which can take from 30 to 60 minutes for the analysis to complete. Solid or liquid samples to be analyzed are normally inserted into small glass capillary tubes which are then inserted into the direct probe tip. Samples analyzed via the direct probe technique are normally restricted to the analysis of pure single component solid samples due to the complexity of the mass spectra if multicomponent samples are analyzed. This limitation of the direct probe technique has been minimized by utilizing the mass spectrometer in the chemical ionization (CI) mode. Mixtures of samples containing 3 to 8 components can be readily analyzed and the components distinguished in the CI mode. Most mass spectrometer direct probes have a temperature limitation of 400 degrees C. This low temperature limits the usefulness of the technique. In many instances it is desirable to analyze samples at much higher temperatures. One may wish to analyze a high molecular weight drug sample which is not volatile at the lower temperatures. In other cases one may wish to cause the thermal decomposition of a compound in order to determine the decomposition products and fragments. A new high temperature direct probe has been designed to permit the analysis of samples at probe temperatures up to 800 degrees C.
Probe Design
Figure 1 - High Temperature Direct Probe and Controller
Figure 1 shows the high temperature direct probe and programmable temperature controller which was designed and manufactured by Scientific Instrument Services for this study. The probe shown was designed for use on a Hewlett-Packard 5989 MS Engine mass spectrometer. Another model was also designed and manufactured for the Finnigan TSQ mass spectrometer. For both of these instruments no modification of the mass spectrometer or probe inlet and valve assembly is required. The probe handle contains the connector for the interconnecting cable, one gas input line for the introduction of cooling gas and an output or exhaust for the cooling gas. The probe tip temperature is controlled via a separate digital temperature controller. A cable connects the probe to the controller. An additional cable connects the temperature controller to the remote start input on the mass spectrometer to permit the remote starting of the mass spectrometer from the probe temperature controller. The temperature controller permits the programmable temperature ramping at ramp rates from 1 to 200 degrees per minute. A detailed description of the design of the high temperature probe tip is shown in Figure 2. The probe tip has been designed such that the sample vial is placed in direct contact with the heater coil. This permits the maximum heat transfer to the sample to enable the high temperatures achieved with this probe and also permits the fast temperature ramp rate obtainable with this probe. The miniature resistance heater coil is surrounded with ceramic insulation and encased in an Inconel sheath. The thermocouple which monitors the probe tip temperature is located at the bottom of the sample vial well. This not only measures the temperature but also provides the signal to control the temperature of the probe tip. This design permits the probe to be heated up to 800 degrees C. In the ballistic mode the probe will heat up at 700 degrees per minute to the set temperature.
Figure 2 - Enlargement of High Temperature Probe Tip
Experimental
All experiments were conducted on a Hewlett Packard 5989A mass spectrometer. The mass spec was used in the electron impact (EI) mode for this study and was scanned from 50 to 1000 daltons. The source temperature was set to 260 degrees C and the quads to 100 degrees C. The high temperature probe was ramped from 25 degrees C to 800 degrees C at a rate of 200 degrees per minute. The mass spec was started at the beginning of the temperature ramp and data was collected for a minimum of eight minutes for each sample. Two high molecular weight samples were analyzed to determine the feasibility of analyzing these samples at the higher probe temperatures. In addition a sample of the high temperature polyimide Vespel (TM) was analyzed to determine the usable temperature limits of this material as well as the nature of the thermal decomposition products at high temperature.
Results and Discussion
The total ion profile of a sample of the drug Cephalexin (MW = 347) is shown in Figure 3. This drug sample was volatilized from the direct probe sample vial into the mass spec source at a temperature between 240 and 600 degrees C. The mass spectrum of the sample at the maximum level as indicated in the total ion profile is shown in Figure 4. The major ion in the mass spectrum is the 303 ion. This ion is due to the loss of 44 from the M+ ion which can be explained by the loss of the COOH group from the molecular ion. A small molecular ion (mass 347) was detected. This sample demonstrates the usefulness of this technique to analyze high molecular weight drug samples without sample decomposition at these high direct probe sample temperatures.
Figure 3 - High Temperature MS Probe Analysis of Cephalexin
Figure 4 - Mass Spec of Cephalexin at Probe Temperature of 400 degrees C
Figures 5 and 6 show the total ion profile and the mass spectrum for the pesticide Avermectin B1A. This compound has a molecular weight of 872. With the direct probe, none of the compound was volatilized into the mass spec below 400 degrees C. The optimum quantity of this compound was volatilized into the mass spec source at a temperature of 480 degrees C. The major peaks observed in the spectra are characteristic of the reported spectra of this pesticide. Once again this high temperature direct probe proves its usefulness in the analysis of high molecular weight compounds that could not be analyzed via conventional low temperature direct probes.
Figure 5 - High Temperature MS Probe Analysis of Avermectin B1A
Figure 6 - Mass Spec of Avermectin B1A Sample at Probe Temperature of 480 degrees C
The high temperature polyimide Vespel is widely used as a manufacturing component in many instruments including mass spectrometers. It is rated for useful temperatures up to 400 degrees C in these instruments with negligible outgassing or material decomposition. As a manufacturer of mass spectrometer parts and accessories we were concerned about the true usable temperature limitations of this polymer and the nature of the thermal emissions or decomposition products at high temperature. A 0.5 milligram (500 ug) sample of finely ground Vespel was analyzed via the high temperature direct probe and the total ion profile is shown in Figure 7. This was raw material as provided by the manufacturer. No preconditioning or baking out of the material was performed prior to analysis. Although some emissions were detectable between 300 and 600 degrees C, the major components eluted or thermally decomposed from the Vespel sample at 740 degrees C. The resulting mass spectrum of the volatile emissions at 740 degrees C is shown in Figure 8. The resulting spectra, is probably a mixture of more than one component due to the apparent complexity of the spectrum. CI techniques could be used to analyze the sample and verify this assumption. The results do indicate the usable temperature range of the Vespel material. Additional studies are needed to verify the identity of the eluted compounds and to determine if the volatiles which were detected at the lower temperatures could be eliminated by preconditioning of the Vespel material before it was analyzed. The above example demonstrates the usefulness of the high temperature probe for the analysis of samples at high temperature to determine their useful temperature range and to determine the thermal decomposition products.
Figure 7 - High Temperature MS Probe Analysis of Vespel
Figure 8 - Mass Spec of Vespel at Probe Temperature of 740 degrees C
Conclusion
A new high temperature direct probe has been designed and manufactured to permit the direct analysis of solid matrix samples at temperatures up to 800 degrees C. This extends the range of compounds that can be analyzed. We have shown the application of this technique for the analysis of high molecular weight compounds which are not volatile below 400 degrees C. Mass spectra are obtainable without sample decomposition. It has also been demonstrated that this probe can be used to effect the thermal decomposition of a solid sample. This was done to determine the useful temperature range of the material as well as study the chemical nature of the decomposition products. The probe has proven useful to analyze a large number of compounds and fill a gap between the standard direct probes and the pyrolysis probes which are used in mass spectrometers.
The use of a direct insertion probe is a common method for the introduction of samples into the mass spectrometer. The technique is preferred due to the speed of analysis of solid matrix samples. Direct probe samples can be analyzed in as little as 2 minutes as compared to GC samples which can take from 30 to 60 minutes for the analysis to complete. Solid or liquid samples to be analyzed are normally inserted into small glass capillary tubes which are then inserted into the direct probe tip. Samples analyzed via the direct probe technique are normally restricted to the analysis of pure single component solid samples due to the complexity of the mass spectra if multicomponent samples are analyzed. This limitation of the direct probe technique has been minimized by utilizing the mass spectrometer in the chemical ionization (CI) mode. Mixtures of samples containing 3 to 8 components can be readily analyzed and the components distinguished in the CI mode. Most mass spectrometer direct probes have a temperature limitation of 400 degrees C. This low temperature limits the usefulness of the technique. In many instances it is desirable to analyze samples at much higher temperatures. One may wish to analyze a high molecular weight drug sample which is not volatile at the lower temperatures. In other cases one may wish to cause the thermal decomposition of a compound in order to determine the decomposition products and fragments. A new high temperature direct probe has been designed to permit the analysis of samples at probe temperatures up to 800 degrees C.
![]() Yttria coated filament at start |
![]() Yttria coated filament after 16,000 cycles |
New Pumps Added to the List of LOW COST S.I.S. Rotary Vane
Vacuum Pumps - Save 35 - 50% off the price of comparable vacuum pumps.
Adixen (Alcatel) ACP Series Compact Dry Roughing Pumps - a frictionless pumping module, optimized for operation without internal lubricant.
- Direct Replacement for the Shimadzu QP5000 Filament.
PCI Analytics Products - product solutions for all Analytical Instrumentation
SilTite FingerTite, the next generation GC capillary ferrule system without the use of any tools.
(25% discount until Oct 31)
Direct Replacement for the Heated Capillary in the Thermo-Fisher Scientific TSQ & LCQ Mass Spectrometers now available
- Ideal for use in analytical instrumentation
- Will improve your laboratory workflow and reporting confidence in many ways
DigiVac Bench Top Vacuum Gauges
- Vacuum instrumentation with everything you need to go to work
Imtakt HPLC Columns
-offer a wide variety of unique stationary phases in their HPLC
columns
Programmable syringe pumps for controlled flow. Model NE-300 priced at only $275.00
NEW: Wiley 9th Edition - 660,000-850,000
spectra (a 65% increase).
NEW: Agilent 5973/75 Replacement Source
Heater and Temperature Sensor
NEW: SIS Direct Replacement
Filament for the Amatek/Dycor RGA Leak Detectors and Quadrupoles
(p/n 95381VE), which tests have shown increase service life.
NEW: SIS
Replacement Filter Elements For Edwards Oil Mist Eliminators -
Save up to 25% off the price of comparable Edwards replacement
elements.
Identify mass spectra via library
search with the fully-evaluated, most widely used mass spectral
reference library, compiled by the National Institute of Standards
and Technology (NIST).
View and capture images or video at
20x/400x power on a PC computer via attached USB cable.