|
SIS AP2000 AutoProbe - Description of SystemNew Automated Direct Insertion Probe for the PolarisQ and TRACE DSQ Mass Spectrometers from Thermo Electron Corp.
The AutoProbe System is an Automated Direct Exposure Probe (DEP) for the Finnigan PolarisQ and TRACE DSQ Mass Spectrometer Systems. The AutoProbe System injects samples directly into the MS source for mass spec sample analysis with no chromatographic separation. This permits samples to be analyzed in 3 minutes or less. The AutoProbe operates in the Thermo Finnigan Xcalibur MS Software environment and is fully integrated with the Xcalibur software, including system setup, methods setup, system operation and data storage.
The direct probe was one of the earliest techniques used to introduce
samples into a mass spectrometer. The technique is still popular
today because it provides a means to perform rapid sample analysis
with minimal or no sample preparation. The electron impact (EI) mass
spectrometer mode of operation can be used when the sample to be
analyzed is relatively pure or when chromatographic separation of a
mixture is not required. Direct probe analysis has also been widely
used with Chemical Ionization (CI) mass spectrometry. The simplicity
of the CI spectrum permits the analysis of mixtures, whose spectra are
difficult to interpret using the EI mode of operation.
Figure 2 - AutoProbe System Components
Figure 3 - AP2000 AutoProbe DEP Probe Tip (Plug in Replaceable Tips) The Direct Exposure Probe (DEP) uses a platinum wire on the end of a probe for the analysis of samples. Samples for analysis are dissolved in a suitable solvent and 0.125 to 1.0 ul of the prepared sample solution is injected onto a small platinum wire loop on the tip of the probe. After the probe has been inserted into the mass spec source, the platinum wire is heated with a small current to desorb the sample into the mass spec source for analysis. This DEP probe filament can be set to a constant current or can be programmed in up to three ramp steps. This versatility of analysis enables the user to optimize the analysis depending on the sample being analyzed.
The DEP or DCI technique is used for the analysis of samples that are
not volatile enough to be introduced into the mass spectrometer via a
GC and therefore must be directly introduced into the mass
spectrometer source. This technique is known as a 'soft' ionization
technique in which thermally labile samples can be ionized with less
fragmentation compared to conventional EI (Electron Impact) mass spec
analysis. A typical DEP analysis can be completed in 1 to 3 minutes.
However, the technique is normally used for single component samples
since minimal separation of samples is possible with this technique
and samples with 2 or more components may be difficult to
interpret.
Figure 4 - Injection of Sample onto the DEP Probe via
the CTC Autosampler syringe A CTC PAL Autosampler is interfaced to the AutoProbe. A wide range of sample trays can be fitted to the CTC PAL Autosampler from which the samples can be loaded onto the DEP probe wire via a syringe. The system permits the cleaning of the syringe between samples. The system also permits the accurate and reproducible delivery of 0.125 to 1.0 ul samples onto the DEP Probe wire. The AutoProbe adds the dimension of automatic sample loading and unattended analysis to the DEP probe technique. Samples can be analyzed at the rate of 2.0 to 5.0 minutes per sample. Sample analysis time can vary, but a typical sample analysis time is composed of the steps listed below:
1. Position Probe for Sample Loading 15 seconds
The above typical analysis above requires 3 minutes per sample. However several of the steps can be increased or decreased, depending on the samples being analyzed.
Figure 5 - Thermo Finnigan Xcalibur Instrument Configuration Window
Figure 5 - AutoProbe Method Setup Window for entering probe operational parameters AutoProbe samples for analysis are placed into sample vial blocks on the CTC PAL Autosampler (Figure 4). A syringe picks up the sample and delivers it to the injection area on the Autoprobe console top. The syringe then injects the sample onto the wire loop at the end of the DEP probe tip.
Figure 6 - Injection of sample onto the DEP wire coil
The standard sequence of operation of the AutoProbe system is as follows.
Figure 7 - AutoProbe Analysis of Caffeine Sample (AP1000 on the TRACE MS) For additional details click onto the detailed descriptive material on the AutoDesorb system below. The application notes below demonstrate the operation of the AutoProbe for the analysis of samples. For additional pricing and ordering information, contact John Manura at Scientific Instrument Services, or call (908) 788-5550 ext: 800. AutoProbe DetailsArticles and Application Notes
© 1996-2008 Scientific Instrument Services, Inc.
[Trademarks]
|