MS TIP No. 18 -
What techniques or methods do you use to determine if the electron
multiplier is in need of replacement? How do you determine multiplier
gain? What is an acceptable signal to noise level?
Mass Spec Tips is a collection of tips relating to the operation
and service of mass spectrometers. Some of these Tips have been
reported in our newsletter, "The Mass Spec Source."
If you have a question relating to the operation or service of
mass spectrometers or would like to add your comments to the below
tip and would like to have it included in this forum, please contact us.
Input to this database of MS Tips is needed from our readers
to complete this section.
Answer 1
by John J. Manura,
Affiliation: Scientific Instrument Services
We posed these questions in our last newsletter because they are
often asked by mass spec users. The determination of the gain
or proper operation of the electron multiplier may be quite difficult
to determine. Factors such as the cleanliness of the mass spec
source, filament operation or alignment and vacuum system performance
can all erroneously convince the operator that the multiplier
is in need of replacement. On many of the early magnetic sector
mass spectrometers, a sample was introduced and the ion beam was
defocused so that only a few ions per second would strike the
multiplier. These single ions were then detected on the oscillographic
recorder and the peak area in coulombs was calculated. From the
charge ratio, the multiplier gain could be determined. This method
was accurate but cumbersome to use. However, today oscillographic
recorders are rare and defocusing of the ion beam to obtain single
ions may be difficult or impossible to achieve on modern quadrupole
mass spectrometers. Most users determine the performance and usefulness
of the multiplier based on experience and how old it is. Normally
if the mass spectrometer is not performing up to specs, and the
multiplier is more than one year old, the operator will replace
the multiplier. On most mass spectrometers which use autotuning
programs to adjust electron multiplier voltage, the tune voltage
will increase with age. When the mass spectrometer begins to tune
in excess of 2800 volts, it is time to replace the multiplier.
More accurate methodology is needed in order to determine multiplier
performance and to eliminate errors in judging multiplier performance
due to the performance of other mass spectrometer components.
This is the reason that we proposed this question to our readers.
The following responses were received from our readers in the
last month. If you have any additional comments or suggestions,
we would like to hear from you.
Answer 2
by: Marianne List
Affiliation: EcoTest Laboratories
From our experience we have noticed a couple of changes that happen:
1) The electron multiplier voltage goes up to 2999 in an autotune.
2) Will not hold tune for very long
3) Source will look like it needs to be cleaned
4) Length of time between changes (approx. 9 months, Galileo)
We use a Hewlett Packard GC 5890, with a Hewlett Packard MS-5970.
Type of samples run are wastewater, soils, and liquid waste.
Answer 3
by:Barry Austern,
Affiliation: U. S. EPA
In the July 1994 issue of the Mass Spec Source you requested when
a multiplier needs replacement. Coincidentally, just yesterday
I wrote a requisition to you to order a new multiplier for my
INCOS-50B.
Older Finnigan instruments had a cup or shield connection on the
back flange, as well as a scope display, and the way to determine
multiplier gain was well laid out in the manual. The newer instruments,
such as the INCOS series, however, do not have this ability and
the multiplier gain must be determined in a round about way. Frankly,
there is no way of getting a numerical value for the gain, but
one can see when the multiplier is getting sick enough such that
it needs to be replaced.
Over time the multiplier will require more and more voltage to
perform it's job. That is the first indication of multiplier wear.
A new multiplier will work fine at 1000 volts and an old one may
need over 2000 to do the same job. Furthermore, it seems, at least
in my experience, that one loses high mass sensitivity more than
low mass sensitivity. For that reason, in order to keep spectra
consistent, I consider 1800-2000 volts my upper limit. If I have
to use more than that to get decent sensitivity it is time to
replace.
Of course, there are several things that can cause a loss of sensitivity,
from a dirty source to something as simple as a leaky septum causing
sample to be lost. The way to convict the multiplier of the crime
is to determine the noise level. All multipliers have noise spikes,
but, like sensitivity, you need to go to higher and higher voltages
to see the noise. In the INCOS-50 series of instruments go to
MTUN and with the multiplier and filament off set the zero to
give a few noise spikes. Turn on the multiplier. Starting at a
low voltage incrementally increase the voltage and see at what
voltage the spiking gets worse. A new multiplier might do that
at as low a voltage as 1400 volts or so. (Use the ES command in
MTUN to set the voltage.) You might have to use the U command
to reset the multipliers upper voltage limit so you can go to
higher levels. In the case of the multiplier in my instrument
now, the one in need of replacement, 2600 volts gave very little
appreciable increase in noise level.
Answer 4
ELQ400 Multiplier Replacement
by Tom O'Hara
Affiliation: Extrel Mass Spectrometry
Before assuming that the multiplier is bad, check the gain by
following the procedure described either in the 400 manual or
in the Service Short from 1/18/91, which deals with performing
a multiplier gain check. If the multiplier appears to be capable
of reaching a gain of 105, this may be enough to allow you to
continue to run samples. However, excessive noise may cause dropouts
to appear in either the total ion chromatogram or the individual
mass chromatograms.
Sources other than the multiplier may be causing the noise. An
ion volume which needs cleaning can greatly reduce the signal
to noise ratio. Just as detrimental, is an air leak of enough
magnitude to impede either the sample flow, or the flow of ions
from the ionizer to the multiplier. Check to see that the chamber
pressure is at a normal level. Typically, for a singularly pumped
system, this pressure should be in the low 10-5 torr range or
below. For a differentially pumped system, the front, or source
chamber, will normally be in the low 10-5 range, while the rear,
or analyzer chamber, will be approximately an order of magnitude
lower. In general, the lower the pressure, the less obstruction
there will be to the ion flow through the quadrupole resulting
in better sensitivity.
Another factor which could make the multiplier appear noisy, if
you are trying to observe peaks produced from the cal gas, is
the amount of cal gas actually being ionized. Check to see that
there is a pressure increase of approximately 1 to 2 to the 10-6
torr in the source chamber when cal gas is turned on. If your
chamber pressure is normally in the 10-5 range, you may have difficulty
determining if this increase occurs. Either turn down the GC column
flow or remove the column completely, capping off the glass interface
with a blank ferrule (included in the spare parts kit). This should
lower the chamber pressure to the 10-6 range at least, making
the cal gas flow easier to observe.
Even though the cal gas flow may be correct, it should be confirmed
that an air leak does not exist, with either the cal gas on or
off. If the pressure changes correctly when the cal gas solenoid
is toggled, but there is more air than cal gas being delivered,
the signal to noise ratio will be low. If a leak is detected,
take a logical approach to resolving it by observing peaks in
the 10 to 40 mass range while squirting methanol on various joints.
By toggling the divert and cal gas solenoids separately, you shouldn't
let the glass interface intimidate you, but be cautious not to
over tighten any of the fittings. Always hold the fitting firmly
with the appropriate size wrench while tightening the nut against
it! if new ferrules are indicated, replace them.
If a GC column is installed, be aware of the flow rate. There
is a limit to how much flow the system can handle; thus the reason
for a jet separator in cases where the flow exceeds 1 to 2 mls./min.
Most GC manufacturers provide a test column with the instrument
which may be only a few meters long. This column is designed to
be used with a flame detector and not a mass spectrometer. Although
the vacuum gauge may indicate a reasonable operating pressure;
within the mid 10-5 range, most vacuum gauge controllers are calibrated
with nitrogen and therefore may display an inaccurate reading.
Excessive carrier gas flow from a short column or one with too
large an I.D., may cause so much background noise that the baseline
on the oscilloscope will actually rise when the multiplier voltage
is increased. Too high of a chamber pressure, which also creates
background noise, will cause this baseline rise to occur as well.
A multiplier that exhibits this symptom under normal conditions
is most likely contaminated and should be replaced.
Answer 5
by: William Schutzer
Affiliation: Oregon State University
We have a HP 5890 GC with a J&W DB5-MS column interfaced to
a 5971A MS running under HP ChemStation software. We determined
that the multiplier (detector) needed replacing from information
directly off the Standard Autotune output and general machine
performance.
The autotune output yields a large amount of information pertinent
to multiplier performance. The values for EMVolts, X-Ray and the
shape of the 502 PFTBA peak are what we used as diagnostics. Originally,
we noticed that the EMVolts started to rise well above our normal
operating range of 1700 to 2000. In fact it reached 3000, and
stayed there. At this point, we assumed that our source was dirty,
so we cleaned the source. This had no effect. Tunes still produced
EMVolts at or near 3000. Our second clue was that the shape of
the 502 PFTBA peak was fragmented and the abundance for the 69
peak was very low; sometimes 70% - 80% of normal. Our third clue
was the value for the X-Ray skyrocketed up to 130.
In conjunction with the values from the tune, was how the machine
was operating. Our quality control abundance's and the signal
to noise ratio both dropped by 50%.
After replacing the multiplier, all problems were solved and the
machine is back to running as it normally does.
Answer
6
by: Richard A. Berger
Affiliation: Washington University Medical School
Multiplier Gain and Mass
Spectrometer Sensitivity
The gain of an electron multiplier can most simply be defined
as the ratio of the output electron current to the input ion current
at a specified second dynode voltage. The output current is not
directly measurable without auxiliary test equipment, but it can
be inferred from the height of an ion in SIM mode. The input ion
current, on the other hand, cannot be measured on most mass spectrometers
short of instrument modification, and the latter is not recommended.
As mentioned by Barry Austern in the September, 1994 issue of
the Mass Spec Source, older machines such as the Finnigan 3200/3300
could be used to measure the input ion current, as they were equipped
with a normally grounded x-ray shield directly on axis with the
quadrupole rods. One could simply turn the multiplier voltage
off and transfer the preamp lead from the multiplier signal output
to the x-ray shield.
Various methods have been proposed over the years to circumvent
the problem of not being able to measure the input ion current,
and thereby determine the multiplier gain. Even if I can determine
the gain of a multiplier, I would not replace a multiplier based
solely on its gain. Rather I base the replacement decision upon
whether or not routine analyses which could be performed in the
past, can still be performed. I stress the word routine since
many factors can effect instrument sensitivity.
In a research facility such as ours there are not many routine
analyses. Therefore I have devised my own tests which are based
on the functional tests performed by the instrument manufacturers
technicians at the point of initial installation. In order to
run the tests, I use Methyl Stearate in Electron Impact (EI) mode
(M+=298) and Positive Chemical Ionization (PCI) mode with Methane
([M+H] + = 299), and Octafluoronaph-thalene in Negative Chemical
Ionization (NCI) mode (M- = 272). I make sure that the tests are
performed with a clean source on a well pumped down machine using
the same column (DB-1, 12 meter, 0.2 mm ID, 0.33 u film), temperatures,
flow rates, and acquisition parameters each time. Then, I test
to determine two things:
1) What is the minimum amount of material I need to inject to
produce a signal to noise ratio of 10:1 when monitoring the molecular
ion (or M+H ion in PCI)? (I inject progressively smaller amounts
of material while varying the multiplier voltage until I determine
this amount/voltage point).
2) By using 1000 times more material and the same multiplier voltage
as in 1 above, what is the relative standard deviation of the
M+1/M, M+2/M and M+3/M ratios? This latter test is important to
a research lab such as ours because much of our work involves
ratio measurement using stable isotopes. The first test may actually
suffice for most labs.
When a multiplier is new the first test yields low amounts of
material at low voltage. Then, as the multiplier ages, both will
rise.
These tests will yield numbers which will provide comparative
ratios over a period of time, and therefore help determine continued
effectiveness of an electron multiplier. However, my ultimate
criteria is still whether or not routine analyses can be performed.
Answer
7
by: Richard Milberg
Affiliation: University of Illinois
All the later VG 70/ZAB's and our MAT CH5 mass spectrometer have
Faraday cups at the end of the ion beam path. The signal can be
measured there and compared to the output of the multiplier. On
older 70/ZAB's and MAT instruments the signal could be measured
at either the first dynode of the multiplier or through the resistor
string with the voltage on.
Another method is based on measuring the signal strength of single
ion events employing an oscilloscope of known display response
and gain. The gain can be measured on any mass spectrometer with
an analog scope display and adjustable frequency response whose
detector pre-amplifier's input impedance is known. It does not
require any special connections or modifications to the detector.
Multipliers must be replaced when they need replacing. Generally
we replace conventional venetian blind or discrete dynode multipliers
when they saturate at 107 gain. Multipliers such as the old Kramer
17-dynode (FK7 or FK17) as well as the EMI 119 can achieve gains
of >109 when new. The gain of the photodetector systems on
the VG 70-SE4F and 70-VSE in the Lab have not appreciably changed
in 7 and 5 years respectively. This is the advantage of a photodetector
system, the EMI photomultipliers are sealed systems and are in
a high vacuum environment as well.
© 1996-2008 Scientific Instrument Services, Inc.
[Trademarks]
|