Mass Spec Tips is a collection of tips relating to the operation
and service of mass spectrometers. Some of these Tips have been
reported in our newsletter, "The Mass Spec Source."
If you have a question relating to the operation or service of
mass spectrometers or would like to add your comments to the below
tip and would like to have it included in this forum, please contact us.
Input to this database of MS Tips is needed from our readers
to complete this section.
Affiliation: Scientific Glass Engineering
The occurrence of peak tailing and poor reproducibility can occur with standard syringe injection of liquid samples into the GC injection port.
This is a common problem which is consistent with slow volatilization of the sample and/or slow transfer of all of the sample onto the column. It occurs with microvolume syringes when the minute sample being delivered from the syringe forms droplets which wet the syringe needle rather than forming a jet of nebulised liquid from the needle tip. As the needle is rapidly withdrawn from the injection port, it is wiped by the septa causing slow secondary vaporisation near the cool septum cap. Not only poor peak shape is observed but sample size reproducibility will be poor.
Position of quartz wool in the injection liner
Resolution of the problem is achieved by effectively wiping the sample from the needle tip during the injection. High speed autosamplers commonly use quartz wool as the liner packing. The position of the quartz wool should be adjusted to achieve effective wiping of the needle.
Theoretically the quartz wool should be placed in the hottest part of the injector, normally the center (Figure #1). In this position, the quartz wool offers no wiping of the needle tip. Figure #1a shows the resultant chromatogram of a 0.3µl injection from a 0.5BR-HP-0.63 syringe into such an injection port. Unacceptable solvent peak tailing and a poor reproducibility (see Table I) results.
0.2µl injection of 0.02% C10 in Heptane chromatographed on a 25QC3/BP1-0.5 column.
Injector 200 degrees C and oven 120 degrees C.
TABLE I - Reproducibility of 15 injections @0.3 µl
Centrally packed liner Liner packed to wipe syringe needle
15 Injections C10 @0.3µl 15 Injections C10 @0.3µl
RSD = 12.73% RSD = 0.57%
By adjusting the position of the quartz wool so that the needle penetrates approximately 1/3rd the way into the quartz wool on injection (Figure #2), the resultant chromatograms show excellent peak shape (Figure #2a) and reproducibility (see Table I). Linearity of dispensed volume is also improved (Figure #2b). The adjustment of the packing position does not alter the boiling point discrimination characteristics of the injection system.
Recommendation
Adjusting the position of the injection point packing so that the needle tip is wiped during injection, improves peak shape, reproducibility and linearity without adversely affecting boiling point discrimination.
Additional Hint:
Do not over tighten the septum cap when using a high speed autosampler. Deformation of the septum may cause leakage during the rapid injection stroke.
![]() Yttria coated filament at start |
![]() Yttria coated filament after 16,000 cycles |
New Pumps Added to the List of LOW COST S.I.S. Rotary Vane
Vacuum Pumps - Save 35 - 50% off the price of comparable vacuum pumps.
Adixen (Alcatel) ACP Series Compact Dry Roughing Pumps - a frictionless pumping module, optimized for operation without internal lubricant.
- Direct Replacement for the Shimadzu QP5000 Filament.
PCI Analytics Products - product solutions for all Analytical Instrumentation
SilTite FingerTite, the next generation GC capillary ferrule system without the use of any tools.
(25% discount until Oct 31)
Direct Replacement for the Heated Capillary in the Thermo-Fisher Scientific TSQ & LCQ Mass Spectrometers now available
- Ideal for use in analytical instrumentation
- Will improve your laboratory workflow and reporting confidence in many ways
DigiVac Bench Top Vacuum Gauges
- Vacuum instrumentation with everything you need to go to work
Imtakt HPLC Columns
-offer a wide variety of unique stationary phases in their HPLC
columns
Programmable syringe pumps for controlled flow. Model NE-300 priced at only $275.00
NEW: Wiley 9th Edition - 660,000-850,000
spectra (a 65% increase).
NEW: Agilent 5973/75 Replacement Source
Heater and Temperature Sensor
NEW: SIS Direct Replacement
Filament for the Amatek/Dycor RGA Leak Detectors and Quadrupoles
(p/n 95381VE), which tests have shown increase service life.
NEW: SIS
Replacement Filter Elements For Edwards Oil Mist Eliminators -
Save up to 25% off the price of comparable Edwards replacement
elements.
Identify mass spectra via library
search with the fully-evaluated, most widely used mass spectral
reference library, compiled by the National Institute of Standards
and Technology (NIST).
View and capture images or video at
20x/400x power on a PC computer via attached USB cable.