C27
Thermal Desorption Low Flow Air Sampling Pump
FAX: 908-806-6631
www.sisweb.com
•
Designed for Air Sampling
thru SIS Desorption Tubes
•
Ultra compact size
•
Ultra low flow 5 - 200 cc/min
•
Back Pressure to 25" of H
2
O
•
LCD clock display
•
Fault Flow Indicator
•
RFI shielding case
•
UL and OSHA approved
•
One year warranty
•
Adapter fitting for SIS
Thermal Desorption Tubes
Features:
Specifications
Operating range
5-200 cc/min, back pressure to 25"
H
2
O
Pressure range
Back Pressure up to 25" H
2
O
Flow control
±5% of set point.
Battery System
Rechargeable, plug in, 4.8V 500 MAH
NiCad battery pack.
Designed to meet
all UL requirements.
Internal charging
(external charging with adapter)
Size
2-1/2" W x 1-3/8" H x 4-5/8" L
Weight
12 ounces
Temperature
Operating:
-20° to 45°C
Storage: -40° to 45°C
Charging:
5° to 45°C
Flow fault indication LED
Battery check LED
Belt clip
Dual filtration system
External flow adjust
Intrinsic safety
The LFS-113 provides for constant flows between 5 to 200 ml/min at variable back pres-
sures up to 25" of H
2
O for use in pumping air or gas samples through the SIS™ Short
Path™ Thermal Desorption Sampling Tubes. Flows are adjustable via an external flow
adjustment, and once sampling begins flow remains constant over variable back pressure
from 0 to greater than 25 inches of H
2
O. An adapter fitting and hose for SPTD tubes and
a battery charger is included with the Air Sampling Pump Kit.
Part No.
Description
914000
Part #914000—Click for details
Current Price: $1364.00
Unit of Measure:
EA
Est. Ship Weight=3 lbs
Not in stock. Standard lead time for out-of-stock items is approximately two weeks.
Model LFS-113 Air Sampling Pump
includes Battery pack, battery charger,
3' of tubing and adapter fitting #781010
Part #781010—Click for details
Current Price: $44.00
Unit of Measure:
EA
Est. Ship Weight=0.05 lbs
Quantity in Stock=4
800086
Part #800086—Click for details
DISCONTINUED
Unit of Measure:
EA
Est. Ship Weight=1 lbs
Replacement Battery Pack
800093
Part #800093—Click for details
DISCONTINUED
Unit of Measure:
PK
Est. Ship Weight=1 lbs
Replacement Filter Kit, Pkg of 3
316TT
Part #316TT—Click for details
DISCONTINUED
Unit of Measure:
Foot
Est. Ship Weight=0.2 lbs
3/16" dia. Tygon™ Tubing
781010
Part #781010—Click for details
Current Price: $44.00
Unit of Measure:
EA
Est. Ship Weight=0.05 lbs
Quantity in Stock=4
Adapter Fitting
Low flow - dual modes
The LFS is an ultra low flow sampler
designed for air sampling on the SIS
Thermal Desorption Sample Tubes.
Constant flow mode sampling
Ideal for single or "in-series" tube sam-
pling providing constant flow rates from 5cc
to 200 cc/min.
The flow is adjusted through
the sampler with an adjust flow screw.
The pump will automatically maintain
the flow set over the constant flow range at
back pressures up to 25" H
2
O.
External flow adjust
Externally adjustable flow control valve,
settable over the full range of the flow.
Flow fault indication LED
A latching fault function is standard on
all LFS-113 samplers.
An LED fault indica-
tor will light if the unit is operated outside its
performance range. If the fault is not cor-
rected within approximately 30 seconds,
the pump will shut off and the LED will
remain lit. The time will lock and display the
time at which the fault occurred, ensuring a
valid sample. If the fault condition is
cleared, the LED will extinguish and the unit
will continue to operate normally.
Battery check LED
When the unit is turned on, an LED indi-
cator signals to show that the battery will
provide a minimum of eight hours service
over the operating range.
Rechargeable battery pack
The LFS has a rechargeable battery
pack system which makes it completely
portable.
Battery packs are recharged via a
standard 120V charger.
Dual filtration system
The dual filtration system provides an
internal pump filter providing a needed
backup to prevent destructive gas and
moisture vapors from entering and damag-
ing the sampler. The inlet filter is easily
unscrewed and filters replaced for preven-
tive maintenance protection.
Intrinsic safety
UL approval for use in hazardous loca-
tions Class I Groups A, B, C, D, Class II
Groups E, F, G, and Class III for explosive
gas and dust atmospheres.
Warranty
The LFS sampler is guaranteed for one
year against defects in parts and/or work-
manship.
Elapsed
time clock
A time elapsed clock provides for accu-
rate timing and valid sampling data. If the
unit should fault, the clock will latch the time
and indicate on the display. When the fault
is cleared the sampler may be re-started
and the clock will begin a new timing
sequence.
RFI shielding case
This sampler is provided with an EMI-
RFI shielded case to dissipate electromag-
netic and radio frequency interference, pro-
tecting the pump electronics and ensuring
valid sampling data. Pumps have been
designed to dissipate EMI between 13.5
MHz and 495 MHz at the exposure level set
by ANSI C95. 1-1982.
Model LFS-113 Sampler
See Page C46 for Applications
- Table of Contents
(Print Catalog)Table of Contents of the Scientific Instrument Services general catalog volume 36. - Print Catalog C. Short Path™ Thermal Desorption SIS Vol. 36 catalog Section C - Short Path™ Thermal Desorption.
- Thermal Desorption Accessories
(Catalog C26) - Thermal Desorption Low Flow Air Sampling Pump
(Catalog C27) - Gilibrator™ Air Flow Calibration System
(Catalog C28)