A new PC-controlled Short Path Thermal Desorption System for the analysis of volatiles and semi-volatiles by GC and GC/MS systems.

Main Features:
A Complete Description of the Theory of Operation is presented in Technical Bulletin No. 1.
The new TD-5 Short Path Thermal Desorption System consists of a Thermal Desorption Unit, Electronics Control Unit, Control Software, and optional Cryo-Trap (see figure above). The Thermal Desorption Unit sits directly on top of the GC injection port of most GC's, where it is utilized for the direct desorption of both volatile and semi-volatile samples into the GC injection port and column. Due to its "short path" of sample flow, the TD-5 overcomes the shortcomings of previous desorption systems by eliminating transfer lines (which can be easily contaminated by samples) and by providing for the optimum delivery (and therefore maximum sensitivity) of samples to the GC injector via the shortest path possible, i.e. direct injection into the GC, like using a GC injection syringe.
When ready to be analyzed, a syringe needle is attached to the desorption tube which is then attached to the connector tube on the Autoinjector Assembly of the Desorption Unit.
Purging: The carrier gas through the Desorption Unit is turned on via the electronics console microprocessor and the flow through the desorption tube is adjusted via the flow controller, rotameter and/or the pressure transducer between 1.0 ml/min and 120 ml/min.
Injecting: The desorption tube is injected with the needle attached. The desorption tube will pass through the opening in the middle plate of the Desorption Unit base to position the desorption tube in proper alignment with the GC injection port and the normally open desorption block assembly.
When injection is complete, the flows are readjusted as required by the method of analysis, i.e. split/splitless, etc. In this position the sample is not being desorbed into the GC since the heating blocks have not yet closed around the desorption tube. The temperature of the tube remains close to room temperature due to the action of the cooling fan, which pulls in room air from the front of the Desorption Unit, thru the view port and across the desorption tube. Carrier gas flows, desorption temperatures, and GC parameters can be adjusted as required.
Blocks close and desorption: The Microprocessor Control actuates an air valve which delivers air to the air powered solenoid and moves the hinged heating blocks from the open to the closed position around the desorption tube. The desorption tube will ballistically heat up to the set temperature or the temperature program program ramp for the heater blocked will begin. The combination of the heat applied and the carrier gas flow through the Desorption Tube will purge the desired components into the GC injection port and onto the front of the GC column.
The various parameters are set and utilized according to the application requirements. Normally desorption temperatures between 70 degrees C and 250 degrees C are suitable for most applications. The maximum desorption temperature permissible with the system is 400 degrees C.
The heater blocks can be temperature programmed at ramp rates up to 100 degrees/min. Normal desorption times vary from 3 minutes to 15 minutes, however longer desorption times up to 100 minutes are permitted. Since the column is normally maintained at subambient temperatures, the desorbed compounds of interest are trapped on the front of the GC column in a narrow band utilizing GC oven cooling or the GC Cryo-Trap Accessory. Despite the long desorption times, the peaks eluted from the column are extremely sharp and well resolved.
The temperature and time parameters are configured in the software.
The TD-5 is operated by control software residing on a PC computer. This software enables the user to configure run parameters (e.g. set-point times and temperatures), initiate runs, and monitor run status (e.g. actual temperatures and pressures). The PC software permits automated operation of the desorption cycle including purging, injecting, desorbing, and remote starting of the GC, as well as control of the GC cryo-trap (cold-trap) accessory.

Figure: Thermal desorption control software showing the
configuration of parameters for a thermal desorption method.
Here, times and temperature set-points are specified.
These are displayed graphically on the right side.

Figure: Thermal desorption control software showing a
run in progress. Shown are the time sequence of the desorption process and
the temperature program (Method parameters), the history of previous
samples run (Sample Log Book), and the status of the current run (times,
temperatures, pressures, and animated graphic of the desorption system).
|
Applications of the SPTD System Thermal Desorption (Purge & Trap) Environmental Air AnalysisResidual Gas, Solvents and Chemicals in: PharmaceuticalsDirect Thermal Extraction of: Plastics see also the Complete List of Application Notes. |
See Ordering Information for the Model TD-5 Short Path Thermal Desorption System
![]() Yttria coated filament at start |
![]() Yttria coated filament after 16,000 cycles |
New Pumps Added to the List of LOW COST S.I.S. Rotary Vane
Vacuum Pumps - Save 35 - 50% off the price of comparable vacuum pumps.
Adixen (Alcatel) ACP Series Compact Dry Roughing Pumps - a frictionless pumping module, optimized for operation without internal lubricant.
- Direct Replacement for the Shimadzu QP5000 Filament.
PCI Analytics Products - product solutions for all Analytical Instrumentation
SilTite FingerTite, the next generation GC capillary ferrule system without the use of any tools.
(25% discount until Oct 31)
Direct Replacement for the Heated Capillary in the Thermo-Fisher Scientific TSQ & LCQ Mass Spectrometers now available
- Ideal for use in analytical instrumentation
- Will improve your laboratory workflow and reporting confidence in many ways
DigiVac Bench Top Vacuum Gauges
- Vacuum instrumentation with everything you need to go to work
Imtakt HPLC Columns
-offer a wide variety of unique stationary phases in their HPLC
columns
Programmable syringe pumps for controlled flow. Model NE-300 priced at only $275.00
NEW: Wiley 9th Edition - 660,000-850,000
spectra (a 65% increase).
NEW: Agilent 5973/75 Replacement Source
Heater and Temperature Sensor
NEW: SIS Direct Replacement
Filament for the Amatek/Dycor RGA Leak Detectors and Quadrupoles
(p/n 95381VE), which tests have shown increase service life.
NEW: SIS
Replacement Filter Elements For Edwards Oil Mist Eliminators -
Save up to 25% off the price of comparable Edwards replacement
elements.
Identify mass spectra via library
search with the fully-evaluated, most widely used mass spectral
reference library, compiled by the National Institute of Standards
and Technology (NIST).
View and capture images or video at
20x/400x power on a PC computer via attached USB cable.